Bug can cause deadly failures when anesthesia device is connected to cell phones

Federal safety officials have issued an urgent warning about software defects in an anesthesia delivery system that can cause life-threatening failures at unexpected times, including when a cellphone or other device is plugged into one of its USB ports.

The ARKON anesthesia delivery system is used in hospitals to deliver oxygen, anesthetic vapor, and nitrous oxide to patients during surgical procedures. It is manufactured by UK-based Spacelabs Healthcare Ltd., which issued a recall in March. A bug in Version 2.0 of the software running on the device is so serious that it could cause severe injury or death, the US Food and Drug Administration warned last week in what's known as a Class I recall. In part, the FDA advisory read:

Reason for Recall: Spacelabs Healthcare is recalling the ARKON Anesthesia System with Version 2.0 Software due to a software defect. This software issue may cause the System to stop working and require manual ventilation of patients. In addition, if a cell phone or other USB device is plugged into one of the four USB ports for charging, this may also cause the System to stop working.

This defect may cause serious adverse health consequences, including hypoxemia and death. Spacelabs Healthcare received one report related to the software defect. There has been no injuries or deaths associated with this malfunction.

At least 16 vulnerable units were in place at hospitals in North Carolina and South Carolina, according to the Class I advisory, the most serious type of recall notice issued by the FDA.

Read 2 remaining paragraphs | Comments